The Model 726 Ambient Ozone Analyzer with Cal/Sample Valves

726 module

 

The Model 726 is a basic analyzer, like the 722, but includes a Cal/Sample valve that can direct either sample gas or calibration gas to the Equivalent Method Photometer. The valve can be switched either by external control via digital inputs or by internal timer control.

Installed in either Docking Station, the Model 726 is an ideal replacement for the Dasibi 1008.

Another version of the 726 is the 722V which is designed to be paired with the Model 724 in a Dual Docking Station, the 722V/724 combination instrument is an ambient ozone analyzer with an integrated ozone calibrator configured in a very compact package!

 

 

Rear Panel

Ozone Photometer Specifications

726 rear

Analog outputs:

4 isolated 16-bit outputs with ±100 mV to ±10V independently selectable ranges

Full Scale Range:

0 - 100 to 10,000 PPB, independently programmable for each output

Sample Flow Rate:

.5 to 2.0 LPM, adjustable

Photometer Cycle Time:

3 to 20 seconds, adjustable

Zero Noise:

< 0.0003 PPM

Span Noise:

< 0.5% of reading (above 100 PPB)

Precision:

< 1 PPB

Linearity:

< 1% of full scale

Zero Drift:

< .001 PPM per month

Span Drift (per 1 wk):

< 1% of reading

Lag / Rise / Fall Times:

10 / 20 / 20 seconds, respectively (with fast cycle setting)

 

Module Specifications:

Power

Standard:

+5, +/-15V & +24V at less than 80 Watts

12V Option:

9-16V at less than 7 amps

Dimensions:

7” wide x 4.9” high x 19.5” long (178 x 124.5 x 495 mm)

Weight:

9 lb (4 kg)

 

 

 

Flow Diagram

 

726 flow

 